6 month old infant s/p AV septal defect repair. 3rd
postoperative day. Underlying rhythm was 3rd degree AV block &
probable sinus node dysfunction. The rhythm strip was printed to evaluate the
temporary pacemaker.
Questions: (Answers are in red-color fonts)
What mode is being used? ….DDD
Comment about Atrial pacing, Atrial sensing and
Ventricular pacing?
a.
Is the pacemaker atrial pacing properly?
(Yes/No) …No (No P wave after atrial pacing spikes “AP”)
b.
Is the pacemaker atrial sensing properly?
(Yes/No) …Yes (When there is a native P wave, no atrial
pacing spike. Atrial sensing “AS” properly)
c.
Is the pacemaker ventricular pacing properly?
(Yes/No) …Yes (QRS complex follows every ventricular
pacing spike “VP”)
No comments: