Monday, February 20, 2012

Answer to Quiz #4: EKG - Pacemaker

Question: What's wrong?



Answer:

Ventricular wire is inserted into atrial outlet of the pacemaker and atrial wire into the ventricular outlet!


Finding in the EKG strip: In each beat, first pacing-spike elicits a QRS complex followed by the second pacing spike that elicits a P wave!